My device needs calibration. What should I do?
Please contact your local service partner. You can find a list of service partners and their contact details here.
Please contact your local service partner. You can find a list of service partners and their contact details here.
There is a known bug in the FTDI USB communication driver version below 2.12.36.3.
Please remove the driver as described in the manufacturer's driver uninstallation guide.
Then download and install the latest version of Mira from the software download page.
Login into Mira Section system setting → user management on instrument (uncheck).
Store settings and transfer the settings to the device via click on Update button.
Download latest version of Mira, which has the latest firmware already included.
After that, connect instrument with PC (USB) and press ‘Update’ button.
The instrument and the PC shall not be disconnected (see display of instrument).
The instrument must have enough energy (full battery) or must be connected to mains.
For more information please refer to the Mira online help.
See also Sentiero Firmware Update Quick Start Guide
Reset the Admin account by starting the Recovery-Console.
This program is installed in the same program folder as Mira.
Password: Stargate.
The password is: 1234
Mira contains an online help function.
Additional information on how to install can be found in the Mira manual (pdf).
Please note that this version is from 2010.
The newer information is available in the online help in MIRA.
Log into MIRA (see above how to do so) to reach to this point.
The system language just influences the language of the login screen.
If you want to change the user-specific user interface language please set the language accordingly in “My Settings”, save the change, and restart the application.
Make sure that "Support RS-232" is selected in the System Configuration tab of Mira.
In rare cases there may be problems with USB3.0 connection.
If possible use a USB2.0 port and retry to connect to the device.
Mira requires a certain operating systems, e.g. for Mira 2.4: Windows 8.1, Windows 10, Windows 11.
Please look up the most recent information at the software download page.
If this happens there can be multiple reasons.
Please check the following:
- Compare the size and SHA-1 Checksum of the Mira setup file with the values given at the website. If differing, please reload the file from the website.
- In Windows Explorer View Settings enable the view of hidden items
- Navigate to "C:\ProgramData" and look for hidden folders which are named like "{A193EEBA-90F2-4921-9FE8-5464100FB332}"
- Locate all folders where a file "Mira.exe" is in one of the subfolders (you may use programs like Mythicsoft's Agent Ransack)
- Delete all folders named like "{A193EEBA-90F2-4921-9FE8-5464100FB332}" from "C:\ProgramData" where there is a "Mira.exe" in a subfolder (e.g. "C:\ProgramData\{A193EEBA-90F2-4921-9FE8-5464100FB332}\OFFLINE\25791C92\F6447A7A")
- Restart the setup
Windows Update sometimes renders the Microsoft SQL Server Compact 3.5 SP2 unusable.
Please do the following steps:
- Navigate in windows to Programs & Features or App & Features
- Uninstall "Microsoft SQL Server Compact 3.5 SP2 ENU" (appears twice with slightly changed naming at 64bit Windows)
- Repeat MIRA installation
This needs to be done before data is loaded to the database of the new computer.
The Mira database is located at: "C:\ProgramData\PATH medical\Mira" for default installations.
It is called "Mira.sdf" (or shown as "Mira" and file type "SQLServerCE_File", depending on the computer settings).
At both computers do a backup of the file.
Then copy the file from the old computer to the new computer.
With the Windows Explorer go to the folder that is shown after Data Source=.
If ProgramData is not shown please set the checkmark at Hidden items in the View menu.
Right click to Mira.sdf or Mira (depending on view settings) and open the Security tab.
Click Edit and change the settings for Users to grant Full Access permissions.
Confirm with OK.