How can I perform hardware tests and what are the possible error messages?
Please refer to the manual, chapter 3.5.4.
Please refer to the manual, chapter 3.5.4.
Please use only the big ear tips together with the big probe tip during measurements and the small ear tips together with the small probe tip. Wrong combination of ear tip and probe tip will deteriorate your results. See also the advice in the accessory box. If in doubt, please perform a probe test as described in the manual, chapter 3.5.4 or contact your distributor.
The display is deactivated after 2 minutes (power safe mode). Please touch the display to restart it. After 5 minutes without activity, the instrument powers down completely. Please press the ‘on-switch’ to restart.
Please contact your local service partner. You can find a list of service partners and their contact details here.
Please restart the device by pressing the reset switch (see fig 3 in the manual). Please charge the battery if necessary. If the error is still present, please contact your distributor.
Please check if there is any dirt/small particle between your display and your housing. Remove the dirt/small particle with a piece of thin paper.
Please check if you can slide the screen left/right to switch between the “most recently used” list and the full list. You may configure the startup default in the settings/preferences dialog.
In order to prevent unintentional changes of the factory defaults / country specific defaults in the U.S.A., the devices’ settings are locked with a password. This password should be told to the supervisor only by the sales representative during the introduction. However sometimes this information is not provided. The password is: MOZART. Please contact your supervisor or sales representative if you need further information.
To do BC audiometry the following prerequisites need to be met: - The “Audio Class 3” licenses is present (see “Activated Modules at Device Settings -> About Device -> Licenses Management) - BC connected at appropriate port: Grey for devices with SN 290000 and below, Red all other devices - Protocol “Expert” selected in the Audio module preset. You may reset the device to factory defaults after applying the license code enabling “Audio Class 3” and use the “Threshold” preset.
When pressing "Details" the device will tell why it complains. If it is because a transducer with passed calibration was connected you may simply delete the list by pressing the "x" button.
Please run a pump unit test. Details of this process are outlined in chapter 3.2.8, section "Practical Use" of the PATH How-To-Manual which can be downloaded from this link.